9. Control Commands
This device operates according to command signals sent from the host computer. Control commands and
data are all sent and received in ASCII codes.
9-A Initialize Commands
The following commands are used to change the status of the device immediately after the power is
turned on:
-1 Kinds of Initialize Commands
9.Control Commands
-2 DLE DL Command (Reading Sensitivity Setting Command)
[Explanation]
This command is for setting the reading sensitivity level.
16 levels corresponding to density levels 1 to 16 can be set.
For example, a mark of density level 2 or more is read when "02" is set.
The setting of the rotary switch is enabled when the power is turned on while software is prioritized
when a DL command is used.
This setting is not stored after the power is turned off.
* This applies to when the CAN command is executed.
In the SR-307 compatible mode, the density levels are 1 to 8.
[Format]
DLE D L N1 N2 CR
(10H) (44H) (4CH) (0DH)
N1N2: 2-byte ASCII code
dnammoCtamroFnoitanalpxE
 LDELD LD〜 .ytisnedgnidaerehtsteS
 DD3CD DD〜
gnidaerehtniecnereffidehtsteS
.ytisned
 T ELD T .tamrofgnimitehtsteS
 3CD
3CD〜
3CD L
.
tuptuoenil-51roenil-21ehtstceleS
ehtnidaerebotsenilehtseificepS
.edomSEroDE,AE,E
 4CD4CD〜 .srehtodnasedocrotarapessteS
 NACNAC RC.ecivedehtsteseR