Model 7000 Series Turbine Flowmeter Section 1
Section1—Introduction
Barton 7000 Series Flowmeters
Barton 7000 Series owmeters are specially designed to accommodate a
broad range of precise uid measurement applications. They can accept larger
ow rates with lower pressure drops than other meters in their class, and fea-
ture a self-ushing design for longer sustained accuracy. Their high-frequency
digital output is ideal for interfacing with digital readout equipment. The 7000
Series' symmetrical bi-directional design supports reverse ow applications,
without reducing accuracy or ow capacity.
All meters are shipped with a pickup coil. Each meter assembly includes two
locknuts, two ow straightener tube assemblies, a center shaft, two diffus-
ers, and a rotor, and is factory-calibrated with water. Bearing type varies with
models: Models 7100 and 7300 feature wear-resistant tungsten carbide sleeve
bearings and thrust washers. Models 7200 and 7400 feature self-lubricating,
precision stainless steel ball bearings. The Model 7400 features bearings with
dry lubricant impregnated ball separators.
During operation, uid passes over the diffuser section and is accelerated onto
a multi-blade hydro-dynamically balanced turbine rotor. The rotor speed is
proportional to the volumetric ow rate. As the rotor turns, a reluctance type
pickup coil (mounted on the meter) senses the passage of each blade tip and
generates a sine wave output with a frequency that is directly proportional to
the ow rate.
The rotor is the only moving part of the Barton turbine owmeter. The small
lightweight rotor hubs ensure fast response to process ow changes. The rotor
is hydro-dynamically balanced during operation, eliminating the need for me-
chanical thrust leveling. This low-friction design improves metering linearity
and reduces wear and maintenance.
The pickup coil is inserted into a conduit adapter that is welded to the outside
of the turbine body. The boss is explosion-proof/ameproof with 3/4" male
NPT electrical threads. The pickup coil can be removed for troubleshooting
without depressurizing the meter or stopping uid ow.
Specications—7000SeriesTurbineFlowmeters
Electrical Classication
• Explosion-proof/Flameproof CSA listed Class I, Division 1, Group
B,C,D; Class II, E,F,G; Class III, Enclosure 4 waterproof to NEC (USA)
and CEC (Canadian) standards, including single seal certication to
ANSI 12.27.0-1