EasyManuals Logo

Setra Systems 5000C User Manual

Default Icon
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
29
ALTERING DISPLAY UPDATE RATE
The INT Mode is factory preset to provide a fast display update (short integration
time) for filling parts to a target count/weight. The use of this mode can be altered to
provide a slow display update (long integration time) for weighing in the presence of
draft or vibration. Follow the procedure below to change the functionality of the INT
mode to either slow or fast display update:
1. Press the — key and enter 11110. This is the INT mode selection access code.
2. Press the setup/cal key. The display will read "INT 1 or 0".
3. Press 1 for a long integration time (vibration control). Press 0 for a short
integration time (filling).
4. Press the setup/cal key.
Note: The integration speed selected will remain programmed even if the scale is off.
Press the ./rcl then conv key to enter the INT mode, and press ./rcl , conv to exit
this mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Setra Systems 5000C and is the answer not in the manual?

Setra Systems 5000C Specifications

General IconGeneral
BrandSetra Systems
Model5000C
CategoryScales
LanguageEnglish

Related product manuals