8
PERCENT DEVIATION
To calculate the amount by which a weight varies from a reference, follow
this procedure:
1. Press TARE.
2. Place the reference weight on the pan.
3. Press the "%" key. After acquiring a stable reading, the display will read
"100.000", "100.00" or "100.0" depending on the amount of weight
applied and the % LED will be lit.
4. Press the TARE key. After acquiring a stable reading, the display will
read "0.000 ", "0.00" or "0.0" depending on the amount of weight
applied. The display now shows percent deviation.
5. Remove the reference weight.
6. Place the weight to be measured on the pan.
7. Wait for the ok LED to light. Read the display. The display indicates
percent deviation from the reference.
8. Remove the weight.
9. Repeat Steps 6-8 as many times as desired.
10. Press UNITS to return to weighing.
NOTE: To display a % of a reference weight, skip step 4.
FILLING TO 100%
You may specify a weight as a given percentage of a total. To fill to 100%,
proceed as follows:
1. Press TARE.
2. Place the reference filled container on the pan.
3. Press the % key. After acquiring a stable reading, the display will read
"100.000", "100.00" or "100.0" depending on the amount of weight
applied.