SI 480E DIN SIZE WEIGHING CONTROLER
5-3-4. Measure the DEAD Weight of Weighing Scale.
dEAd is displayed. Please press key with empty scale.
Then the indicator starts to measure and find optimal Dead weight value of Scale automatically.
It takes 10sec or 20sec to get the best
situation.
Over than 1/10,000 resolution setting,
To guarantee the preciseness, dead weight calculation will be operated twice.
5-3-5. Input Max. Output ( Rated Output Voltage / mV)
Input the output value load cell
Following fixed decimal point.
After mV displayed, Check the Rated output value of Load cell.
(Refer to the load cell label, or Test Report) . And Press key to save and move to next step.
dead
Cal10
1.98700
rnU