23
SI 480E DIN SIZE WEIGHING CONTROLER
4
Activated within 50% of Max. Capacity
5
Activated within 100% of Max. Capacity
6
There is no limit of Zero key operation range.
CAUTION : If setting over than 10%, The display weight could be over than Load cell
input signal or Max. Capacity and it may display CELL-Err or incorrect weight value.
Tare key Operation Range selection : (-) value is same to (+)
F09
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
2
Activated within 50% of Max. Capacity
3
Activated within 100% of Max. Capacity
“Hold” Mode selection
F10
0
Sample Hold : Hold current weight until “Hold Reset
1
Peak Hold : Measure Max. weight value and hold on display.
2
Average Hold : Hold average value (Refer F-F50)
STEADY condition check time setting
F11
03
00
09
During the set time period, estimate weighing parts STEADY condition
and display. If you set small value, indicator will take STEADY fast, if you
set value, indicator will take STEADY slow.
( 0.5sec per set value)
Display Up-Date speed setting
F12
1
60/sec
2
30/sec
3
20/sec
4
15/sec
5
10/sec
6
6/sec
7
3/sec
8
2/sec
9
1/sec
Weight Display selection under Unpass / OverLoadcondition
F13
0
Not Display Weight (just UNPASS or “–OL- is displayed)
1
Display Weight (with a flash)