17 | P a g e
INSTALLATION (CONTINUED)
LIFTING AND HANDLING
The BACTRON is heavy. Use appropriate powered lifting devices. Follow these guidelines when
lifting and handling the BACTRON workstation:
Lift the BACTRON only from its bottom surface.
Doors, handles, and knobs are not adequate for lifting or stabilization.
Restrain the BACTRON completely while lifting or transporting so it cannot tip.
Remove all removeable components, such as shelf spacers and trays, and secure all doors
in the closed position during transfer to prevent shifting and damage.
Note: To prevent damage when moving the BACTRON, turn each of the four leveling feet
completely clockwise.
LEVELING
The BACTRON must be level and stable for safe operation. Each BACTRON ships with four
leveling feet.
1. Insert one leveling foot into each of the four holes in the bottom corners of the workstation.
2. Adjust each foot at each corner until the workstation stands level and solid without rocking.
a. To raise a foot, turn it in a counterclockwise direction; to lower a foot, turn it in a
clockwise direction.
INSTALL THE BACTRON
Install the unit in a workspace location that meets the criteria discussed in the previous entries of
the Installation section.
Do not connect the unit to its power source at this time.