EasyManuals Logo

Shimadzu GC-8A Series User Manual

Shimadzu GC-8A Series
13 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
7
GC-8AIT
Column Oven
Temperature range
Temperature control
Temperature control accuracy
Temperature
Readout (Monitor)
Heating speed
Cooling speed
Column
Overheat Protection
: Room temperature to 399°C
: Proportional-integration type of zero cross
switching system
: ±0.1°C
: When the temperatures of the column
oven and the injection port/detector have
reached the set points, the respective
READY lamps are lit.
:
From room temperature to 350°C in 13 minutes.
: From 350°C to 100°C in 3.5 minutes.
: Stainless steel: 6m x 2
Glass: 5.4m x 2
: Two-stage protection:
(1) If the temperature exceeds the
programmed (preset) temperature by 50°C,
the heater will be automatically turned off.
(2) If the temperature reaches about 420°C,
the heater will be automatically turned off.
Temperature of Injection Port/Detector
Temperature range
Temperature control
Temperature control accuracy
Temperature
Readout (Monitor)
Overheat Protection
Injection Port
: Room temperature to 400°C (10°C steps)
: Proportional-integration type of zero cross
switching system
: ±0.1°C
: When the temperatures of the column
oven and the injection port/detector have
reached the set points, the respective
READY lamps are lit.
: If the temperature reaches about 420°C,
the heater will be automatically turned off.
: On-column injection ports
Flow Control
Carrier gas : Two pressure regulators
Two column inlet pressure gauges
Thermal Conductivity Detector (TCD)
Detector
Maximum temperature
Bridge current
Filament protection circuit
: Differential type of semi-diffusion flow system
Four rhenium-tungsten filaments (100Ω each)
: 400°C
: Constant current system (OFF, 60, 70, 80,
90, 100, 120, 140, 160, 180, 200mA)
: Incorporated
Others
Dimensions
Weight
Power Requirements
Others
: 440W x 405D x 435Hmm
: 26.5kg
: AC100/115V or 220V as ordered.
1500VA max. 50/60Hz.
: (1) Column oven heating starts after the
injection port / detector reaches the set
temperature.
(2) Cannot be used with a preparative system.