TOCV-CP_IMshort_Control-ver-2_Rev07F Page 22 of 22
(Software version ≥ 2.00) effective date: September 2007
4 Editing the Sample Table during analysis
The sample table can be edited during analysis by changing to the Edit Mode. This mode
allows the following operations to be conducted in the sample table in the same way as when
analysis is not being conducted
- During analysis, select “Instrument” – Edit Mode”
- The sample table is placed in the Edit Mode and following procedure can be conducted:
- Adding and deleting samples
- Changing analysis parameters and vial numbers
- Printing reports for samples that are already analysed
- Click [Start] button to cancel the Edit Mode and restart analysis
5 Stop of Measurement
This option interrupts processing of the current injection, and
analysis proceeds to the next injection measurement
This option stops analysis after all of the scheduled injections of the
current samples have been analysed