Menu Operation
cotton swab. Using the cotton swab, start at the center of the lens and
move the swab in a circular motion until you spiral out to the edge of the
lens surface.
c) The lens surface should be clean and free of streaks or smudges.
d) Turn on the power and make sure no smudges or streaks are visible on
the monitor screen. Press X/Y key to change the screen and check the
state of the lens surface on both the X- and Y-screens. Perform dust
check.
⚫ Do not hit or touch tip of electrode when cleaning it or remove electrode
before cleaning objective lens.
⚫ Recommends to clean the objective lens when replace electrodes.
B. Rotating Cleaver Blade
If cleaving quality drops (usually cleave 1000 to 1500 times), the blade
may be worn. Please follow the step to adjust the blade angle: do the
following:
a) Within 1.5mm hex wrench to loosen the compression screw of blade;
b) Rotate blade to the next scale number (generally there are 16 or 24 two
types of blade);
c) Hold the side-face of blade and relock the screw to complete the
adjustment.
C. Blade cycle use
A blade after 16 (24) scales have been used, if adjust the blade height, can
be use again by cycling once or twice:
a) Within 1.5mm hex wrench to loosen two compression screws on the tool
holder;
b) Follow the direction of dial to adjust the height-direction adjusting stud
at the bottom of slider. Turn it as clockwise direction. The white spot on
stud should correspond to the grid of one scale turn;
c) Relock these two compression screws, test cleaving, and check the