Menu Operation
In AUTO mode, certain parameters cannot be changed.
Parameter
Description
Fiber Type
List of splice modes stored in database is displayed. A
selected splice mode stored in the database area is copied to a
selected splice mode in the user-programmable area.
Mode Title1
Title for a splice mode expressed in up to 10 characters.
Mode Title2
Detail explanation for a splice mode expressed in up to 10
characters. Title2 is displayed at the [Splice Mode Select]
menu.
Cleave Limit
Set cleave limit. An error message is displayed if the cleave
angle of either the left or right fiber ends exceeds the selected
threshold.
Loss Limit
An error message is displayed if the estimated splice loss
exceeds selected threshold (loss limit).
Arc Power
In SM/DS/MM/NZ/AUTO modes, the Arc Power is fixed at
40 bits.
Arc Time
Arc Time is fixed at 1500 ms for SM and DS modes, 2000 ms
for NZ mode, and 3000 ms for MM mode. This is
automatically set depending on the fiber type when AUTO
mode is selected.
Cleaning Arc
A cleaning arc burns out micro dust on the surface of the fiber
with an arc discharge for a short period of time. The duration
of the cleaning arc can be changed by this parameter.
Rearc Time
Splice loss may be improved by an additional “rearc”
discharge in some cases. The duration of this additional arc
can be changed by thisparameter.
2. Heater Mode
There are 20 user-programmable heating modes. Select one best suitable for the
protection sleeve used.