Menu Operation
cleaving quality. Follow the steps above until the height of stud is
perfect (to see the trouble removal for details).
D. Replace the blade
When the blade circulation is used, and you cannot improve the cleaving quality
by changing the angle of blade and adjusting the height of stud, the blade needs
to be replaced.
a) Within 1.5mm hex wrench to remove the screw and gasket of blade;
b) Turn blade to the tool holder side, then hold side faces of blade and take
it out;
c) Replace the used blade by a new one, and then screw the gasket and
screw on.
⚫ DO NOT let the blade to collide anything!
Menu Operation
Press button to enter splicer menu, there are six main menus: “Splice Mode
Menu”, “Heat Mode Menu”, “ Calibration and Maintenance Menu”, “Data”,
“Splice Setting Menu”, and “Setting Menu” as shown below: