ADM-860C 04/14/089
nnRc nnnn
This message indicates that backpressure compensated air flow readings are being stored. The display
shows the current reading and its order in the reading sequence during a STORE process.
nn
Ru nnnn
This message indicates that nonbackpressure compensated air flow readings are being stored. The display
shows the current reading and its order in the reading sequence during a STORE process.
nn
S n.nnnn
Displays a reading in memory and its place in the stored reading sequence. Accessed using forward or
reverse RCL keys. Number of decimal places will vary.
nn
S ERASED
This message indicates that the current reading displayed in RECALL (nn
s
) has been erased.
nnS NP 0
This message is displayed when a stored negative pitot tube reading is accessed using a RECALL key.
nn
Sc nnnn
This message indicates a backpressure compensated air flow reading is being viewed in memory. The
display shows the reading and its order in the reading sequence during a STORE process. Accessed using
forward RCL keys.
nn
Su nnnn
This message indicates a nonbackpressure compensated air flow reading is being viewed in memory. The
display shows the reading and its place in the reading sequence during a STORE process. Accessed using
forward RCL keys.
nnu nnnn
Displays the number of nonbackpressure compensated readings in memory and the average of just the
nonbackpressure compensated readings taken during a FlowHood reading sequence.
nn
0 n.nnnn
Displays the number of readings in memory and the average of the readings. Displayed when the STORE
key is pressed during a STORE sequence. Number of decimal places will vary.
3 n.nnnn
Displays the sum of the readings in memory during a STORE sequence. Displayed when the STORE key
is pressed twice if readings are in memory. Number of decimal places will vary.
3c nnnn
Displays the sum of the backpressure compensated readings in memory during a STORE sequence.
Displayed when the STORE key is pressed if readings are in memory. Number of decimal places will vary.
3c NONE
This display indicates that there are no backpressure compensated readings in memory.
3u nnnn
Displays the sum of the nonbackpressure compensated readings in memory during a STORE sequence.
Displayed when the STORE key is pressed if readings are in memory. Number of decimal places will vary.
xx
' n.nnnn
This display indicates that the back-light is on and battery power draw has increased. xx indicates the units
for the reading.