Siemens US15-101.812.01.02.02 ACUSON S2000
05.10 Ultrasound
System Delivery Procedure 15
Page 15 of 40
3. Verify the following LED states:
a) Verify the Power Button flashes green.
Fig. 8: S2000 Control Panel Power Button
b) Verify the lower-rear panel AC OK LED illuminates green. Reference
(3/Fig.7/p.14).
c) Verify the lower-rear panel Battery LED (below the AC OK LED) illuminates green.
Reference (2/Fig.7/p.14).
4. Press the Power button. Reference (Fig.8/p.15).
5. Verify that the following occurs within the first four seconds of the boot sequence:
a) The lower-rear panel LEDs 1,2,3,4,5, and 6 illuminate green. Reference
(4/Fig.7/p.14)