16 System Delivery Procedure
ACUSON S2000 US15-101.812.01.02.02 Siemens
05.10 Ultrasound
Page 16 of 40
b) Verify the Power Button illuminates green. Reference (Fig.8/p.15).
c) The control panel turns on, then off, indicating the completion of the power-on self
test (POST).
d) The I/O panel:
LEDs 1,3, 4, 5, 6, 7, 8, and 9 - illuminate green.
LED 2 - illuminates yellow.
Fig. 9: IO Panel -- LED Lights
6. Verify system completes the boot sequence to imaging without displaying any error
messages.
NOTE
The boot sequence completes in approximately three to five
minutes.