Wire the modified Y cable as follows:
¹ Plug the Y cable's D31.X2 plug into the new D31.X2 board.
¹ Plug the Y cable's D31.X5 plug into the new D31.X5 board.
¹ Plug the Y cable's D1.X14 plug into the D1.X14 board.
¹ Plug the Y cable's X21 plug into the open X21 socket on the footswitch mounting
jack.
Attach the ends of the Y cable along the existing cable assemblies using the cable ties
included in the accessory kit.
Insert the plug for the multi-function footswitch into the footswitch socket and lock it in
place.
Closing the covers 0
Reattach the ground wire to the rear cover panel of the basic unit. Make sure the wire
has good contact with the cover panel.