4 General information and safety guidelines
ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG
09.06 CS PS SP
Page 4 of 20
Medical Solutions
1- 1 General infor mation and sa fety guideline s
General information 0
Validity of the document 0
This document is valid for the subsequent installation of a multi-function footswitch in an
ARCADIS Varic system or an ARCADIS Orbic system that has been prepared for the
multi-function footswitch option.
Requirements 0
The ARCADIS Varic and ARCADIS Orbic systems delivered after December, 2006
(approximately) have already been prepared for the installation of the multi-function foot-
switch. The following are indications that a system has been prepared for this installation:
¹ There is an externally accessible footswitch plug connector socket on the basic
unit.
¹ The holes for attaching the footswitch holder to the cover panel have already
been drilled and are fitted with dummy covers.
¹ The internal wiring for the footswitch plug connection has been prepared for the
upgrade to the multi-function footswitch.
¹ Software version VB13C or higher is installed on the system.
NOTE
It is not possible to upgrade ARCADIS Varic or ARCADIS Orbic
systems with the multi-function footswitch option upgrade kit
unless these preparations have been made.
Required material 0
• Multi-function footswitch option, material number 08079548
Tools and auxiliary equipment required 0
• Standard tool kit
• Cable ties (included in the delivery volume of the multi-function footswitch)