22 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 22 of 34
Medical Solutions
8. Remove the cover from the single tank.
9. Adjust the chamber gain by pushing the up or down button for at least 200 ms for each
push at the chamber (1/Fig.5/p.22).
¹ Use a small Allen key (diameter 0.5 - 1.0 mm) to push the “UP” or DOWN” button
(e. g. 20 x push down = -10%).
¹ “DOWN” = 0.5 % reduction of gain per push.
¹ “UP” = 0.5 % increase of gain per push.
10. Reinstall the single tank cover, check the dose rate again and perform a “Wellhöfer -
chamber test”.
NOTE
A “Wellhöfer chamber test” is necessary after a “Wellhöfer cham-
ber gain adjustment”.
Wellhöfer - chamber test
1. Click on <Go> to start the chamber test.
¹ Status bar: success
¹ Wait about 3 seconds for the constant readout
2. Note the displayed “DAP” or “Air Kerma” from the ARCADIS monitor depending on the
configuration (path: <Configuration>-<Imaging System>-<Display Settings>).
¹ With “DAP” configured, the monitor indication is “ xx.xx cGcm
2
”.
¹ With “Air Kerma” configured, the monitor indication is “ xxxx.x mGy ”.
3. Click on <Go> again to stop the chamber test.
¹ Status bar: success
¹ Window appears: Calibration is finished --> confirm with <OK>.
¹ The main system restarts.
Fig. 5: Wellhöfer_DAP

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals