3-14
123
4
5
ABCDEFG
X041E
Siemens-Elema AB SPB7-230.051.07 Page 14 of 18 M1000/3000 Nova
Solna Sweden Rev. 02 03.00 SPS-UD Wiring Diagrams
D809
D810
LED
X80
X81X82X83
TILT
8
0°, ±10°
D805
D810**
17
TILT
+24VF
X81
X82
X83
X856
X860
2
4
9
6
5
4
3
2
X80
1
X868
+24VF
TILT
V8
S896**
TUBE ANGLE MEASUREMENT, POT. RETURN, TILT SIGNAL
POT_RETURN
POT_RETURN
4
2
5
D801 TESTPOINTS AND LED’s
X813
X816
PROM
CPU
0VA1
7V
5V_REF
0VD2
5V_DSP
X814
X811
F1
0VD1
5V
V7
V6
V5
S1
RESET SWITCH
X855 X868
D805
D803
OPCOM
FORCE
THICKNESS
OK
ANGLE
X816
1-20
X801
1-20
CPU
A/D
1
1
+5V_REF
POT_RETURN
0VA
0VD
0VA
20
+5V_REF
X814
1
5
18
2
17
9
15
X
8
1
3
6
3
0V_M
+
5
V
_
M
X813 35 36
X861 X862
D806
M3
+/- 10°
RD
YE
0V
0V_M
X862
1
2
X861
1
3
4
ROT_POS
D806*
X874
1
PRES_FORCE
THICKNESS
R861
10k
R863
2k
3
5
2POT_RETURN
(SEE PAGE 2-3)
COMPRESSION
UNIT
D805
10k
R871
R803
2k
TUBE ANGLE
PRESET ANGLE
PRESET_ANGLE
TUBE_ANGLE
+5V_REF
X863
3
2
1
6
7
5
X868
8
X855
6
10
1
3
D801
X863
RD
BK
WH
BN
BU
YE
5V_REF
0VA 0VA
ROT_POS
X856
X860
31
1
ROT_POS
* D806 not used with
software v2.1 or higher
+5V
X809
** Not fitted in M1000