Siemens Elema AB Register 6 SPB7-230.051.07 Page 7 of 14 M1000/3000 Nova
Solna, Sweden Rev. 02 03.00 SPS-UD Wiring Diagrams
X041E 4-7
LIST OF SWITCHES AND COMPONENTS, GENERATOR
Component Overview (Generator)/Bauteileübersicht (Generator) 0
Component Page Origin Description
K1 5-6/3F D705 Small/large focus changeover
Umschalten kleiner/großer Fokus
K2 5-8/2D D705 Inverter frequency changeover
Umschalten der Wechselrichterfrequenz
K3 5-8/3D D705 Control for brakes
Ansteuerung für Bremsen
K1 5-7/3D D706 MO/W bias voltage changeover
K2 5-7/3D D706 Connects bias voltage to tube
K3 5-7/4E D706 MO/W filament current changeover for large focus
K4 5-7/2E D706 Bias voltage monitoring
K5 5-7/4E D706 MO/W filament current changeover for small focus
K6 5-7/1D D706 Relay output for xray indication
K1 5-9/3E D710 SS relay
SS-Relais
K2 5-9/1C D711 By-passes intermediate circuit charging resistors
Überbrückt Zwischenkreis Ladewiderstände
K3 5-3/2E D711 Switches on generator and stand
Schaltet Generator und Stativ ein
K4 5-9/3F D711 Starting rotary anode
Anlauf Drehanode
K5 5-9/3F D711 Starting and braking, rotary anode
Anlauf und Bremsen Drehanode
K6 5-3/2E D711 System ON
System EIN
K7 5-3/2E D711 System supply OFF
System AUS
K8 5-6/3C D711 Filament intermediate circuit ON
Heizungs-Zwischenkreis EIN
K9 5-6/3C D711 By-passes charging resistors of the heating
Überbrückt Ladewiderstände Heizung
K10 5-3/2E D711 Switches on P24V
Schaltet Trafo P24V ein
K11 5-9/1B D711 Intermediate circuit ON
Zwischenkreis EIN
K12 5-9/1B D711 Intermediate circuit ON
Zwischenkreis EIN