MAMMOMAT 3000 SPB7-230.091.10 Page 2 of 4 Siemens-Elema AB
Service Rev. 01 02.00 SPS-UD Solna, Sweden
2 - 2 Procedure
6.3. Loosen and remove screws and washers 1 and 2 that hold the compression
unit base beam (Fig. 2).
6.4. Loosen screw 3 (Fig. 2) by half to tilt the compression unit slightly forward.
6.5. Pull connector X872 out of printed circuit board D805.
6.6. Pull out connectors X873 and X874, which are located behind the
compression unit.
6.7. Loosen and remove screw 3 (Fig. 2), and remove the compression unit.
7. Loosen bellow by removing plastic guide blocks inside the bellow.
8. Loosen the screws for the ring (1/Fig. 3).
9. Loosen the rotation stop screw (2/Fig. 4). Then turn the object table arm further
clockwise to make the two screws accessible.
MAM00225
Screw 1
Screw 2
Screw 3
Fig. 2 Removal of compression unit
MAM00692
1
Fig. 3 Position of two of the four screws for the ring