MOBILETT Plus HP SPR8-220.091.01 Page 2 of 2 Siemens-Elema AB
Service Rev. 03 03.03 SPS-UD Solna, Sweden
1 - 2 Prerequisites
Safety information 1
Read all safety notes in the document Service Instructions and pay special attention to
the notes below.
Never work with the system open if the batteries are still con-
nected and the capacitor is charged. If the batteries are con-
nected, the complete system is powered on!
The capacitors in M10 may still be charged even if the system is
switched OFF and the line voltage cable is disconnected. Life-
threatening electric shock hazard exists.
The capacitors in M10 must be considered charged until the pro-
tective measures listed in the document Service Instructions have
been performed.
Be sure to follow the ESD guidelines.
WARNUNGWARNING
CAUTION