Installation 2 - 21
Siemens AG SPR8-220.031.04 Page 21 of 30 MOBILETT Plus/ -E/ -HP
Medical Solutions Rev. 04 09.03 CS PS 24
12. Mount the power supply box as shown in Fig. 25 using 5 mm screws (MC6S
5x16mm) and nuts. Be sure to use washers (KB 5.1mm, included in the kit).
13. Fasten the cable to the new power supply unit (pin plugs K24 and K25) with a
cable tie (2.5x100mm) through the hole (pos. 1 / Fig. 25).
14. The cable must also be fastened with cable ties (2.5x100mm) on the power supply
unit (pos. 1 / Fig. 26).
15. Fasten the other cable shielding connector with a screw to the new power supply
box (pos. 2 / Fig. 26). Remember to place a washer (KB 5.1mm) between the
connector and the power supply box.
MOB00524
Fig. 25 Mount of new power supply
1
Power supply
MOB00589
Fig. 26 Fastening cable to new power supply
1
2