MOBILETT Plus/ -E/ -HP SPR8-220.031.04 Page 22 of 30 Siemens AG
Rev. 04 09.03 CS PS 24 Medical Solutions
2 - 22 Installation
16. Remount the aluminum plate with board D201 and the power supply unit.
Reconnect K201, K202 and K203 to board D201.
17. Replace the cable ties as shown in Fig. 23 to fasten the old cables and the new
cable.
18. The end of the cable with pin plug K23 is placed as shown in Fig. 27. The cable
runs on top of the protective cover over board D8 (pos. 1 / Fig. 27) along with the
other cables.
19. Fasten the end with a cable tie (2.5x100mm) to one of the cables already present
(see pos. 1 / Fig. 28). Leave about 7 cm hanging.
MOB00525
Fig. 27 Cable from power supply
1
D8
MOB00526
Fig. 28 Fastening of cable
1
approx. 7 c
D8