MultiRanger 200 and Flow Configuration AG040303 10 of 10
APPLICATION GUIDE
Section 7.1 BS3680/ISO 1438/1 V Notch Weir Configuration Table
Parameter number Description to explain the value Configuration Priority
P600 Set to 7 (BS 3680/ISO 1438/1 V Notch Weir) Must be configured.
P602 = (1 & 2) Enter the values of  in degrees in index 1. In
index 2, view the value of discharge coefficient C
e
Must be configured.
P603 Enter the value of maximum head in unit as set in
P005.
Must be configured.
P605 Zero Head Off-set Must be configured because the zero of the weir
is above the channel zero.
P608 Flow rate units Must be configured.
P607 Flow rate decimal Optional, configured to display the flow rate (Q).
Factory set to 2.
P620 Low flow cut-off. Optional, to avoid totalizing at flows below this
set level.
P621 Auto zero head calibration Optional, configured if displayed head has
constant off-set.
Note: Both Primary devices BS 3680 / ISO 4359 Rectangular Flume and the BS 3680 / ISO 1438/1 Thin Plate V notch weir
can also be configured for Ratiometric calculation in case the flow rate units are not supported in P608.