Area dose product measurement device 5
Siemens AG SPR2-240.814.04 Page 5 of 10 POWERMOBIL/ARCOSKOP
Medical Engineering Rev. 02 07.00 TD SD 24
Calibrating the area dose product measuring device 1
CONSTANT parameters = T
w
1
Read and record the T
w
value shown on the electronics unit of the area dose product
measuring device.
Programming 1
Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
Start the service program.
Select the "Adjustment, Parameter..." menu.
Select "Diamentor" in the "Parameter Groups" combo box.
Click on "Get from Unit".
The parameter "1. Diamentor Existing" must be on YES.
Select parameter "2. Constant T
w
.
Enter the parameter T
w
in the "Value Actual:" field.
Click on "Set Value".
Click on "Put to Unit".
Click on "Save to File".
Fig. 1
This adjustment is used to calibrate the area dose product mea-
suring device to the POWERMOBIL/ ARCOSKOP display.
You must exit the service program with "Logoff" and "Quit" in
order to save the modified parameters to diskette as a new file.
Potentiometer
NOTICE
NOTICE