Area dose product measurement device 5
Siemens AG SPR2-240.814.04 Page 5 of 10 POWERMOBIL/ARCOSKOP
Medical Engineering Rev. 02 07.00 TD SD 24
Calibrating the area dose product measuring device 1
CONSTANT parameters = T
w
1
• Read and record the T
w
value shown on the electronics unit of the area dose product
measuring device.
Programming 1
• Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
• Start the service program.
• Select the "Adjustment, Parameter..." menu.
• Select "Diamentor" in the "Parameter Groups" combo box.
• Click on "Get from Unit".
• The parameter "1. Diamentor Existing" must be on YES.
• Select parameter "2. Constant T
w
.
• Enter the parameter T
w
in the "Value Actual:" field.
• Click on "Set Value".
• Click on "Put to Unit".
• Click on "Save to File".
Fig. 1
This adjustment is used to calibrate the area dose product mea-
suring device to the POWERMOBIL/ ARCOSKOP display.
You must exit the service program with "Logoff" and "Quit" in
order to save the modified parameters to diskette as a new file.
Potentiometer