6 Area dose product measurement device
Checking the programmed value 1
Prerequisites
• The T
w
value is already programmed (see Programming).
• Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
• Start the service program. Select the "Adjustment, Calibrations..." menu.
• Answer the query with "Yes".
• Select "Diamentor" in the "Function Group:" combo box.
• In the "Available Functions" window select the line "Start Diamentor Test" and click on
"Execute".
• Press the >0< key (reset key for fluoroscopy time). The current value is displayed on the
monitor.
• The tested T
w
value is displayed on the monitor of the POWERMOBIL/ ARCOSKOP.
• This value must match the programmed CONSTANT parameter T
w
. Tolerance: ± 5 %.
• If the ->0<- key is pressed (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
• The potentiometer (Fig. 2) of the area dose product measuring device can be used to
adjust the test value
T
w
to the programmed value T
w
.
• After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
• The test is concluded.
• Check the accuracy of the area dose product measuring device (see next page).
Fig. 2
Checking the programmed values will activate the self-test in the
electronics unit of the area dose product measuring device.
This serves to recalibrate the electronics unit.
Potentiometer