POWERMOBIL/ARCOSKOP SPR2-240.814.04 Page 6 of 10 Siemens AG
Rev. 02 07.00 TD SD 24 Medical Engineering
6 Area dose product measurement device
Checking the programmed value 1
Prerequisites
The T
w
value is already programmed (see Programming).
Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
Start the service program. Select the "Adjustment, Calibrations..." menu.
Answer the query with "Yes".
Select "Diamentor" in the "Function Group:" combo box.
In the "Available Functions" window select the line "Start Diamentor Test" and click on
"Execute".
Press the >0< key (reset key for fluoroscopy time). The current value is displayed on the
monitor.
The tested T
w
value is displayed on the monitor of the POWERMOBIL/ ARCOSKOP.
This value must match the programmed CONSTANT parameter T
w
. Tolerance: ± 5 %.
If the ->0<- key is pressed (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
The potentiometer (Fig. 2) of the area dose product measuring device can be used to
adjust the test value
T
w
to the programmed value T
w
.
After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
The test is concluded.
Check the accuracy of the area dose product measuring device (see next page).
Fig. 2
Checking the programmed values will activate the self-test in the
electronics unit of the area dose product measuring device.
This serves to recalibrate the electronics unit.
Potentiometer
NOTICE