Replacement of the 15" LCD Color TFT Display 15
Siemens AG SPR2-130.841.03 Page 15 of 20 Siremobil Compact L
Medical Solutions Rev. 02 06.05 CS PS 24
Manual programming 1
Legitimation 1
• Open the OSD menu.
• Scroll to the menu "Options" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Scroll to the sub menu "Legitimation" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Scroll to the line "Code A:" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Set the value "49" with the "Up "or "Down" keys
and confirm the value by pressing the "Enter" key.
• Scroll to the line "Code B:" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Set the value "60" with the "Up" or "Down" keys
and confirm the value by pressing the "Enter" key.
• Scroll to line "back" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
Interlaced mode 1
• Scroll to the menu "Presentation" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Scroll to the line "Interlaced" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Set the parameter to "AUTO" with the "UP" or "Down" keys
and confirm the value by pressing the "Enter" key.
• Scroll to line "back" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Scroll to line "back" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
The main menu is displayed.
In the normal case, the correct mode will be detected
automatically.
If the previos checks and adjustments could be carried out
successfully, skip this section.
NOTE