Siremobil Compact L SPR2-130.841.03 Page 18 of 20 Siemens AG
Rev. 02 06.05 CS PS 24 Medical Solutions
18 Replacement of the 15" LCD Color TFT Display
Horizontal and vertical position 1
• Scroll to the menu "Mode Adjustment" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Scroll to the line "Horizontal Position" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Set the parameter to "175" for 50Hz video frequency or to "160" for 60Hz video
frequency.
• If necessary change the value with the "UP" and "DOWN" keys, till the horizontal position
of the displayed test pattern is as shown in Fig. 6.
• Confirm the adjusted value by pressing the "Enter" key.
• Scroll to the line "Vertical Position" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
• Set the parameter to "64" for 50Hz video frequency or to "44" for 60Hz video frequency.
• If necessary change the value with the "UP" and "DOWN" keys, till the vertical position of
the displayed test pattern is as shown in Fig. 6.
• Confirm the adjusted value by pressing the "Enter" key.
• Scroll to line "back" with the "UP" and "DOWN" key
and select it by pressing the "Enter" key.
The main menu is displayed.
Fig. 6