Replacement of the 15" LCD Color TFT Display 9
Siemens AG SPR2-130.841.03 Page 9 of 20 Siremobil Compact L
Medical Solutions Rev. 02 06.05 CS PS 24
Replacing the switched-mode power supply 1
• Disconnect the low voltage plug from the TFT display.
• Disconnect the power plug of the switched-mode power supply.
• Loosen the cable ties of the low voltage cable and pull it back to the log book
compartment.
• Take out the switched-mode power supply from the log book compartment.
• Install the new switched-mode power supply in the log book compartment, using new
cable ties.
• Install the low voltage cable to the TFT display and connect it.
• Fix the low voltage cable with cable ties.
• Connect the power plug to the switched-mode power supply.
• Check the function of the TFT display.
The switched-mode power supplies are installed in the log book
compartment and are fixed with cable ties.
Fig. 2
NOTE