EasyManuals Logo

Siemens SIREMOBIL Iso-C Installation Instructions

Siemens SIREMOBIL Iso-C
6 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2 background imageLoading...
Page #2 background image
SIREMOBIL SPR2-130.812.06 Page 2 of 6 Siemens AG
System Manual Rev. 01 12.02 CS PS 24 Medical Solutions
2
Installation and Startup1
Safety information 1
Validity of this document 1
This document is applicable if a key switch is to be retrofitted to the SIREMOBIL Compact,
SIREMOBIL Compact L, SIREMOBIL Iso-C, POWERMOBIL, and ARCOSKOP systems.
Documents required 1
System manual of the product
Parts required 1
Tools required 1
Tool case
1 set of Allen keys
1 drilling machine
1 vacuum cleaner
Measuring equipment required 1
Avoidance of light to severe injuries which may even lead to
death and avoidance of material damage.
The product-specific safety information given in the system man-
ual as well as the safety information contained in the ARTD,
part 2, must be observed.
After all work has been completed and all covers have been
attached, perform the protective conductor test according to
ARTD-002.731.17. The protective conductor resistance must not
exceed 0.2 Ohm.
Retrofit kit key switch 07140895
Protective conductor resistance meter
e. g. "Safety Tester Bender UNIMET 1000 ST", Mat. No. 51 38 727
WARNING

Other manuals for Siemens SIREMOBIL Iso-C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIREMOBIL Iso-C and is the answer not in the manual?

Siemens SIREMOBIL Iso-C Specifications

General IconGeneral
BrandSiemens
ModelSIREMOBIL Iso-C
CategoryMedical Equipment
LanguageEnglish

Related product manuals