SIREMOBIL SPR2-130.812.06 Page 6 of 6 Siemens AG
System Manual Rev. 01 12.02 CS PS 24 Medical Solutions
6 Installation and Startup
Installing the paneling 1
• Reattach the protective plate behind the keyboard in the logbook compartment.
Take care that the cables are not squeezed.
• Refit the paneling of the logbook compartment. Make sure the ground wire connections
are correct when doing this.
• Reinstall the rear paneling of the monitor trolley. Make sure the ground wire connections
are correct when doing this.
Final work 1
• After all work has been completed and all covers have been attached perform the protec-
tive conductor test according to ARTD-002.731.17... The protective conductor resis-
tance must not exceed 0.2 Ohm.
• Add these instructions to the system manual of the product.
• File the Addendum/Supplement to the Operating Instructions included in the scope of
supply into the existing operating instructions.
Customer information 1
• Demonstrate the function of the key switch.
• Hand over the keys.
• Hand over the operating instructions.