3 - 1
Siemens AG SPR2-230.812.02 Page 1 of 2 SIREMOBIL
Medical Solutions Rev. 02 02.06 CS PS SP System Manual
Settings 3
TFT display BARCO MVG 1318, part number 80 79 530 3
• Make the checks and if necessary settings according to the replacement instructions
TD00-000.841.17...
Refer to the System-dependent settings chapter of the instructions.
Concluding work 3
• Close all covers again. Make sure that the protective ground wire connections make
good contact.
• Perform the protective ground wire test according to TD00-000.860.01.
The protective ground wire resistance must not exceed 0.2 ohms.
• Perform the IQ quick test of the system.
• Perform the country-specific acceptance tests (part acceptance).