2 - 2 Installation of TFT displays
• Unscrew the two bases of the monitors from the monitor trolley and remove them com-
pletely (Fig. 5).
• Remove the two video cables.
• Completely remove and disconnect the two power cables.
• 4 prepunched holes are visible around the central cable bush (Ê/ Fig. 5).
4 metal bolts are inserted in these holes and they are thus closed.
Drive out the 4 metal bolts downwards with a drift (or another round object). Remove the
4 metal bolts from the monitor trolley.
• Lift off the central blind cover and close one of the cable bushes on the side with it
(Fig. 6).
• Close off the 2nd cable bush on the side and the 8 fastening holes around the cable
bushes on the sides with the enclosed blind covers.
Fig. 5 Monitor attachments Fig. 6 Monitor trolley cable bushes