SIREMOBIL SPR2-230.812.02 Page 2 of 8 Siemens AG
System Manual Rev. 02 02.06 CS PS SP Medical Solutions
2 - 2 Installation of TFT displays
Unscrew the two bases of the monitors from the monitor trolley and remove them com-
pletely (Fig. 5).
Remove the two video cables.
Completely remove and disconnect the two power cables.
4 prepunched holes are visible around the central cable bush (Ê/ Fig. 5).
4 metal bolts are inserted in these holes and they are thus closed.
Drive out the 4 metal bolts downwards with a drift (or another round object). Remove the
4 metal bolts from the monitor trolley.
Lift off the central blind cover and close one of the cable bushes on the side with it
(Fig. 6).
Close off the 2nd cable bush on the side and the 8 fastening holes around the cable
bushes on the sides with the enclosed blind covers.
Fig. 5 Monitor attachments Fig. 6 Monitor trolley cable bushes
Mark for the keyboard tray