The cleaning unit can be set at:
Automatic cleaning
Set the relay output of the signal converter at cleaning.
Commissioning
Automatic cleaning
Set the required cleaning cycle time. The default value is 24 hours. Use the lock key to lock the set values.
Return to the operator menu using the top-up key. The signal converter is now set at automatic cleaning.
Technical data
Cleaning voltage
AC cleaning
DC cleaning
60 V AC
30 V DC
Cleaning period 60 s + 60 s pause period
Relay
• Load 42 V AC / 2 A; 24 V DC / 1A
Operation
• Automatic
• Manual
Switch relay activated when cleaning is in progress
Yes
No
Indicator lamps LEDs: “ON” and “CLEANING”
Supply voltage and power consumption 115 ... 230 V AC, +10% ... -15%, 50 ... 60 Hz, 7 VA cleaning,
5 VA stand by
11 ... 30 V DC / 11 ... 24 V A C, 50 ... 60 Hz, 7 VA cleaning,
5 VA stand by
Ambiens temperature
• During operation
• During storage
-20 ... +50°C (-4 ... +122°F)
-20 ... +70°C (-4 ... +158°F)
Electrical noise (EMC)
Enclosure
• Material
• Width
• Height
• Rating
• Mechanical load
Standard 19” insert in aluminium/steel (DIN 41494)
21 TE (4.75”)
3 HE (5.25”)
IP20 / NEMA 2 to EN 60529 and DIN 40050
1 g, 1 ... 800 Hz sinusoidal in all directions to EN 60068-2-36
Dimensions
Weight 1 kg
CENELEC Emission Emmunity
Light industry EN 50081-1
Heavy industry EN 50082-2
We have reviewed the contents of this publication to ensure consistency with the hardware and software described. Since variance cannot be precluded entirely, we cannot guarantee full con-
sistency. However, the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions. Responsibility for suitability and intended use of
this instrument rests solely with the user.
Copyright © Siemens AG 05.2008 All Rights Reserved. Technical data subject to change without prior notice.
Siemens Flow Instruments A/S
6430 NORDBORG
DENMARK