MAG 8000 & MAG 8000 CT
24
SFIDK.PS.026.D9.02
FT/PDM Meter Display Parameter/data type Factory settings Data range
number version view Fix parameter or meter data that not are changeable
312 All - Flow direction totalizer 2 Reverse forward, reverse or bi-directional net flow
313 All - Totalizer 2 changes date PS3 production date and time
320 All - Verification mode enable No Yes / No
321 All - Calibration date Calibration date year-month-day T
hours:minutes:seconds
323 All - Calibration factor Sensor related
324 All - Gain correction Sensor related
325 All - Sensor offset Sensor related
327 All - Adjustment Factor 1 -2 to 2
328 All - Low flow cut off 0.05% 0 to 9.9%
329 All - Filter time constant 5 Tau 1 to 1000
331 All - Excitation frequency limit 6.25Hz for advanced version 1/15Hz, 1/5Hz, 1.5625Hz, 3.125Hz,
and 1/15Hz for basic version 6.25Hz, 1/30Hz, 1/60Hz
332 All - Excitation frequency sensor limit Sensor related 6.25 Hz (DN25...DN200 (1"...8"))
3.125 Hz (DN250..DN600 (10"..24"))
333 All - Empty pipe detection enable Yes Yes / No
334 All - Empty pipe limit 25000 ohm = 20uS/cm 0 to 2.15*10^9
400 All - Output A enable Yes Yes / No
401 All Se3 Pulse A function Forward Forward, Reverse, Forward net,
Reverse net
402 All Se3 Amount per pulse A Sensor related 0-1*10^10
403 All - Pulse width for pulse A 50 ms 5 ms, 10 ms, 50 ms, 100 ms, 500 ms
404 All - Output B enable Yes Yes / No
405 All Se4 Pulse B function Alarm pulse, alarm, call-up
406 All - Pulse B direction Reverse Forward, Reverse, Forward net,
Reverse net
407 All Se4 Amount per pulse B Sensor related 0-1*10^10
408 All - Pulse width for pulse B Sensor related 5 ms, 10 ms, 50 ms, 100 ms, 500 ms
420 All M5 Device Communication Address 1 1 to 32
421 All M6 Baudrate 19200 1200, 2400, 4800, 9600, 19200, 38400
422 All M7 Parity Even 1 stop Even 1 stop, Odd 1 stop, None 2 stop,
None 1 stop
423 All - Interframe space 35 35 to 255
424 All - Response delay 5 1 to 50 ms
425 All - Reset communication driver No Yes / No
500 All - Latest service date PS3 production date and time year-month-day T
hours:minutes:seconds
501 All - Operating hours since power up 0 hours
502 All - Battery operating time 0 hours
505 All - Power supply Power supply level Battery or mains power
506 All - Numbers of power up 0
507 All - Battery power MLFB depended 1 to 4 batteries
508 All - Battery change enable No Yes / No
509 All Se1 Battery installation date PS3 production date and time year-month-day T
hours:minutes:seconds
510 All Se2 Actual battery capacity 100% 100 to 0%
512 All - Excitations no. 0
513 All - Power status 0 0: Normal operation, 1: Battery alarm.
Actual battery capacity is below battery
alarm level (% of max capacity), 2: Too
low power (enters stand by mode), 3: As
value 1 and 2 together, 4: External power
gone, 5: As value 1 and 4 together, 6: As
value 2 and 4 together, 7: As value 1 and
2 and 4 together
514 All - Transmitter temperature Actual degree celsius
540 All - Electrode impedance A Measured values 0-185000 ohm
541 All - Electrode impedance B Measured values 0-185000 ohm
542 All - Low medium impedance alarm 0 0 to 2.15*10^9
550 All - Coil current disable No Yes / No
551 All - Fixed flow mode enable No Yes / No
552 All - Fixed flow value 0 -1*10*10^9 to 1*10*10^9
553 All - Flow alarm limit 1000000000 0 to 1*10*9
560 All - Repair checksum No Yes / No