EasyManua.ls Logo

Siemens SITRANS F M MAG 8000 - Page 51

Siemens SITRANS F M MAG 8000
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAG 8000 & MAG 8000 CT
51
SFIDK.PS.026.D9.02
MAG 8000 CT MI-001 verified and labeled products at a given Q3 and Q3/Q4 = 1.25 and Q2/
Q1 = 6.3 measurering ranges se below table:
DN 50 65 80 100 125 150 200 250 300
(2“) (2½“) (3“) (4“) (5“) (6“) (8“) (10“) (12“)
„R“ Q3/Q1 25 25 25 25 25 25 25 25 25
Q4 [m
3
/h] 19 31 50 75 125 188 313 500 750
Q3 [m
3
/h] 15 25 40 60 100 150 250 400 600
Q2 [m
3
/h] 3.78 6.30 10.1 15.1 25.2 37.8 63.0 100.8 151.2
Q1 [m
3
/h] 0.60 1.00 1.60 2.40 4.0 6.0 10.0 16.0 24.0
DN 50 65 80 100 125 150 200 250 300
(2“) (2½“) (3“) (4“) (5“) (6“) (8“) (10“) (12“)
„R“ Q3/Q1 63 63 63 63 63 63 63 63 63
Q4 [m
3
/h] 19 31 50 75 125 188 313 500 750
Q3 [m
3
/h] 15 25 40 60 100 150 250 400 600
Q2 [m
3
/h] 1.50 2.50 4.0 6.0 10.0 15.0 25.0 40.0 60.0
Q1 [m
3
/h] 0.24 0.40 0.63 0.95 1.59 2.38 3.97 6.35 9.52
DN 50 65 80 100 125 150 200 250 300
(2“) (2½“) (3“) (4“) (5“) (6“) (8“) (10“) (12“)
„R“ Q3/Q1 63 63 63 63 63 63 63 63 63
Q4 [m
3
/h] 19 31 50 75 125 188 313 500 750
Q3 [m
3
/h] 15 25 40 60 100 150 250 400 600
Q2 [m
3
/h] 1.50 2.50 4.0 6.0 10.0 15.0 25.0 40.0 60.0
Q1 [m
3
/h] 0.24 0.40 0.63 0.95 1.59 2.38 3.97 6.35 9.52
The label is placed on the side of the encapsulation. An example of the product label is shown
below:
Note
Installation conditions for MAG 8000 CT has to be mounted in Integral (compact) and horizontal
position only, to obtain the MI-001 certification.
Battery packs must be installed with the top part in upwards direction to reach maximum capacity.
8. Technical data