6 Service and maintenance
6.1 Basic safety instructions
Note
The device is maintenance-free.
6.1.1 Maintenance
The device is maintenance-free, however, a periodic inspection according to pertinent directives and regulations must be
carried out.
An inspection can include check of:
Ambient conditions
Seal integrity of the process connections, cable entries, and cover screws
Reliability of power supply, lightning protection, and grounds
WARNING
Impermissible repair and maintenance of the device
Repair and maintenance must be carried out by Siemens authorized personnel only.
WARNING
Impermissible repair of explosion protected devices
Danger of explosion in areas subject to explosion hazard.
Repair must be carried out by Siemens authorized personnel only.
WARNING
Dust layers above 5 mm
Danger of explosion in hazardous areas. Device may overheat du to dust build up.
Remove any dust layers in excess of 5 mm.
CAUTION
Penetration of moisture into the device
Device damage.
Make sure when carrying out cleaning and maintenance work that no moisture penetrates the inside of the device.
CAUTION
Releasing key lock
Improper modification of parameters could influence process safety.
Make sure that only authorized personnel may cancel the key locking of devices for safety-related applications.
6.2 Cleaning
Clean the outside of the enclosure and the display window using a cloth moistened with water or a mild detergent.
Do not use aggressive cleaning agents or solvents. Plastic components or painted surfaces could be damaged.