EasyManua.ls Logo

Siemens SITRANS LR110 - Measurement Setup - Flow

Siemens SITRANS LR110
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installing/mounting
4.4Measurement setup - Flow
Figure4.8 Agitators
Foam generation
Through the action of filling, stirring and other processes in the vessel, foams which
considerably damp the emitted signals may form on the medium surface.
If foams lead to measurement errors, you should use transmitters with guided radar.
4.4 Measurement setup - Flow
Mounting
In general, the following must be observed while mounting the device:
Mounting the transmitter on the upstream or inlet side
Installation in the centre of the flume and vertical to the liquid surface
Distance to the overfall orifice or Venturi flume
Distance to the max. height of the orifice or flume for optimum accuracy:
>250mm (9.843in)
1
Requirements from approvals for flow measurement, e.g. MCERTS
Flume
Predefined curves:
A flow measurement with these standard curves is very easy to set up, as no
dimensional information of the flume is required.
Palmer-Bowlus flume (Q = kxh
1.86
)
1
The value given takes into account the block distance. At smaller distances, the measuring accuracy is reduced, see ""Technical
data".
SITRANS LR110, HART
Operating Instructions, 10/2023, A5E49829006
17

Related product manuals