Edition 10/2021
Operating Instructions
SITRANS L
Radar Transmitters
SITRANS LR100
https://www.siemens.com/processautomation

Questions and Answers

  • J
    Justin BishopSep 9, 2025
    Why is the 4 … 20 mA signal missing on my Siemens SITRANS LR100?
    • D
      David TuckerSep 9, 2025
      If you observe that the 4 … 20 mA signal is missing on your Siemens Transmitter, it could be due to several reasons: a faulty electrical connection, a missing voltage supply, or an operating voltage that's too low with a load resistance that's too high. First, check the connection and correct if necessary. Next, inspect the cables for breaks and repair if needed. Finally, check and adapt the operating voltage and load resistance if necessary.
  • J
    jared84Aug 23, 2025
    How to fix 'F013 no measured value available' on Siemens SITRANS LR100?
    • J
      jason23Aug 23, 2025
      If your Siemens Transmitter displays 'F013 no measured value available', it indicates there's no measured value during the switch-on phase or operation. You should check and correct the installation or parameter settings, and also clean the antenna system.
  • A
    Anita PalmerAug 25, 2025
    What to do if current signal is greater than 22 mA or less than 3.6 mA on Siemens SITRANS LR100 Transmitter?
    • J
      Jennifer HarveyAug 26, 2025
      If the Siemens Transmitter current signal is greater than 22 mA or less than 3.6 mA, it suggests the transmitter electronics are defective. Depending on the device version, replace the device or send it in for repair.
  • P
    Philip VegaAug 27, 2025
    What does 'M507 Error in the transmitter settings' mean on Siemens SITRANS LR100?
    • M
      matthew92Aug 27, 2025
      If the Siemens Transmitter displays the error 'M507 Error in the transmitter settings', it indicates an error during setup, when carrying out a reset, or auto false echo suppression is faulty. To resolve this, carry out a reset and repeat the setup.
  • B
    breanna72Aug 27, 2025
    How to solve 'F261 Error in the transmitter settings' on Siemens Transmitter?
    • S
      Sara KempAug 28, 2025
      If the Siemens Transmitter shows the error 'F261 Error in the transmitter settings', it indicates an error during setup, auto false echo suppression is faulty or an error occurred when carrying out a reset. You should repeat the setup or carry out a reset.
  • S
    Susan PattersonAug 29, 2025
    How to troubleshoot 'F025 Error in the linearization table' on Siemens Transmitter?
    • H
      Heather AndersonAug 29, 2025
      If the Siemens Transmitter displays 'F025 Error in the linearization table', it's due to index markers not continuously rising, or illogical value pairs. To fix this, check the linearization table, then delete it or create a new one.
  • C
    crawfordsharonSep 18, 2025
    How to fix unstable 4 … 20 mA signal on Siemens Transmitter?
    • J
      James BakerSep 18, 2025
      If the Siemens Transmitter's 4 … 20 mA signal is not stable, it is caused by a fluctuating measured value. To fix this, set damping.
  • F
    Frederick BrownSep 22, 2025
    What does 'F017 Adjustment span too small' mean on Siemens SITRANS LR100 Transmitter?
    • A
      Adam GuzmanSep 22, 2025
      If the Siemens Transmitter displays 'F017 Adjustment span too small' it means the adjustment is not within specification. Change the adjustment according to the limit values (difference between min. and max. ? 10 mm).
  • S
    Shane YoungSep 8, 2025
    How to fix 'F036 No operable software' on Siemens Transmitter?
    • J
      Joan HarrisSep 8, 2025
      If the Siemens Transmitter displays 'F036 No operable software', it indicates a checksum error due to a failed or aborted software update. You can repeat the software update or send the transmitter for repair.
  • J
    Jennifer WilsonSep 4, 2025
    What does 'F040 Error in the electronics' mean on Siemens Transmitter?
    • P
      Paul BallSep 5, 2025
      If the Siemens Transmitter displays 'F040 Error in the electronics', it indicates that a limit value has been exceeded in signal processing or there is a hardware error. Try restarting the transmitter or send the transmitter for repair.

Summary

Preface

Safety instructions for Ex areas

Highlights specific safety precautions for explosive environments.

Safety notes

Authorised personnel

Specifies requirements for individuals operating the device.

Appropriate use

Defines the intended application and conditions for the transmitter.

Warning about incorrect use

Alerts users to hazards from misuse.

General safety instructions

Lists essential safety guidelines for operation.

Installing;mounting

General instructions

Provides guidelines for ambient and process conditions.

Mounting instructions

Gives specific steps and considerations for installation.

Measurement setup - Flow

Explains how to configure the device for flow measurement.

Connecting

Preparing the connection

Details safety and requirements before electrical connection.

Wiring plan

Shows wire assignments for the connection cable.

Access protection

Setup with smart device (Bluetooth)

Diagnostics and troubleshooting

Rectify faults

Explains how to identify and fix malfunctions.

Diagnosis, fault messages

Details how to interpret error codes and messages.

Status messages according to NE 107

Describes device status indicators.

Service and maintenance

Technical data and dimensions

Technical data

Presents detailed specifications of the device.

Siemens SITRANS LR100 Specifications

General IconGeneral
TypeRadar Level Transmitter
Frequency80 GHz
Process PressureUp to 3 bar
Output Signal4 to 20 mA
Output4 to 20 mA / HART
Communication ProtocolHART
CommunicationHART
Enclosure RatingIP68
Measuring PrincipleFMCW Radar
Measuring range0.08 to 30 m
Process Temperature-40 to +80 °C
Power Supply12 to 30 V DC
Housing MaterialAluminum