36 A5E03434626-03, 06/2013
Gauge pressure input, with flush-mounted diaphragm
HART
PROFIBUS PA and FOUNDATION Fieldbus
Measured variable
Gauge pressure
Span (fully adjustable) or
measuring range, max.
operating pressure and
max. test pressure
Span Maximum
operating
pressure
MAWP (PS)
Maximum
test
pressure
Measuring
range
Maximum
operating
pressure
Maximum
test pressure
0.01 … 1 bar g
(0.15 …
14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 … 4 bar g
(0.58 …
58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
0.16 …
16 bar g (2.3
… 232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
16 bar g
(232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
0.6 … 63 bar g
(9.1 …
914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
63 bar g
(914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
Absolute pressure input, with flush-mounted diaphragm
HART
PROFIBUS PA and FOUNDATION Fieldbus
Measured variable
Absolute pressure
Span (fully adjustable) or
measuring range, max.
operating pressure and
max. test pressure
Span Maximum
operating
pressure
MAWP (PS)
Maximum test
pressure
Measuring
range
Maximum
operating
pressure
Maximum test
pressure
43 …
1300 mbar a
(17 ...
525 inH
2
O)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
1.3 bar a
(18.9 psi a)
2.6 bar a
(37.7 psi a)
10 bar a
(145 psi a)
160 …
5000 mbar a
(2.32 …
72.5 psi a)
10 bar a
(145 psi a)
30 bar a
(435 psi a)
5 bar a
(72.5 psi a)
10 bar a
(145 psi a)
30 bar a
(435 psi a)
1 … 30 bar a
(14.5 …
435 psi a)
45 bar a
(653 psi a)
100 bar a
(1450 psi a)
30 bar a
(435 psi a)
45 bar a
(653 psi a)
100 bar a
(1450 psi a)
Depending on the process connection, the
span may differ from these values
Depending on the process connection, the
measuring range may differ from these values
DS III input with PMC connection
HART
PROFIBUS PA and FOUNDATION
Fieldbus
Measured variable
Gauge pressure
Span (fully adjustable) or
measuring range, max.
operating pressure and
max. test pressure
Span Maximum
operating
pressure
MAWP (PS)
Maximum
test
pressure
Measuring
range
Maximum
operating
pressure
Maximum
test pressure
0.01 … 1 bar g
(0.15
14.5 psi g)
1)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
1)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 … 4 bar g
(0.58 …
58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)