Loading...
NOTICE
Use of line and cable entries made of plastic in hazardous areas
Device damage caused by impact at temperatures below -20 °C.
Make sure that the line and cable entries are protected from impacts.
3.2 Installing SITRANS TH in a connection head
Note
The transmitter is only designed for installation in a type B connection head or larger.
The transmitter is either secured in the base of the connection head or in the raised cover of
the connection head.
The scope of the delivery of the transmitter includes:
Springs
Fixing screws
Lock washers for installation on the round plate (SITRANS TH)
Fastening the transmitter:
In the cover of the connection head In the base of the connection head

Transmitter Ceramic base of the measuring element
Connection head
Installing/mounting
3.2 Installing SITRANS TH in a connection head
SITRANS TH320/TH420/TR320/TR420/TF320/TF420
Compact Operating Instructions, 03/2018, A5E41865021-AA 23