EasyManua.ls Logo

Siemens ST-H - 5 Technical Data; Specifications

Siemens ST-H
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Transducers
Compact Operating Instructions, 01/2018, A5E43390688_AA
37
Technical data
5
Note
Device specifications
Siemens makes every attempt to ensure the accuracy of these specifications but reserves
the right to change them at any time.
For a complete listing, including Approvals, see the full operating instructions.
5.1
Specifications
Specification
XRS-5/5C
ST-H
XPS-10
XPS-15/15F
XPS-30
Process Application
Application
Liquids
Liquids and Solids
Max. range
1)
8 m (26 ft)
10 m (33 ft)
15 m (50 ft)
30 m (100 ft)
Min. range 0.3 m (1 ft) Standard:
0.3 m (1 ft)
XPS-15F:
0.45 m (1.5 ft)
0.6 m (2 ft)
Vessel Pres-
sure
Vented to atmosphere
Operation
Beam Angle
10°
12°
Frequency
44 kHz
30 kHz
Temperature
Sensor
Internal
Supply Source
Transducer shall only be supplied by a Siemens certified controller
Environmental
Location
Indoor/outdoor
Max. altitude
2000 m
Max. tempera-
ture
65 °C (149 °F) ATEX:
60 °C (140 °F)
95 °C (203 °F)
Min. tempera-
ture
-20 °C (-4 °F) -20 °C (-5 °F) -40 °C (-40
°F)
Standard:
-40 °C (-40 °F)
XPS-15F:
-20 °C (-4 °F)
-40 °C (-40 °F)
Pollution de-
gree
4
Construction