Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS
08.05 CS PS 24
Urodynamic interface 13
Page 13 of 18
Medical Solutions
• Set Out H-Centre to 79 (press the + or the - key).
• With the Menu key, select Exit and use the + to return to the Advanced menu.
• Press the Menu key until Manual Set appears.
• Press the + key (the sub-menu is displayed).
-Select VGA Left and set 18, until only the X-ray image is visible (press the + or the -
key, default setting is 9).
• Use the Menu key to select VGA Store for a save operation and press the + key until an
acoustic signal is emitted.
• To save all settings, press the Menu key and the + key at the same time..
• Switch the UROSKOP Access off and back on again.
Cutting off the menu bar (FLUOROSPOT Compact) 0
NOTE
This subsection refers exclusively to UROSKOP-Access systems
with the FLC imaging system.
• Press the Menu key until Adjust: Screen size appears.
• Press the + key.
- The sub-menu is displayed.
• Set Out H-Centre to 65 (as setting for PAL; setting for NTSC is 66) (press the + or
the - key).
• With the Menu key, select Exit and use the + key to return to the Advanced menu.
• Press the Menu key until Manual Set appears.
• Press the + key (the sub-menu is displayed).
-Select VGA Left and set 1, until only the X-ray image is visible (press the + or the -
key).
• Use the Menu key to select VGA Store for a save operation and press the + key until an
acoustic signal is emitted.
• To save all settings, press the Menu key and the + key at the same time..
• Switch the UROSKOP Access off and back on again.
Scan converter parameter table (ASPIA imaging system) 0
Tab. 1 Scan converter parameter table (ASPIA )
Parameters Default value (PAL) Default value (NTSC)
Menu“Screen Size“
1
Out H-Centre 79 79
Out H-Width 93 94
Out V-Centre 160 132