Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS
08.05 CS PS 24
Urodynamic interface 15
Page 15 of 18
Medical Solutions
Scan converter parameter table (FLUOROSPOT Compact imaging system) 0
Tab. 2 Scan converter parameter table (FLC)
Parameters Default value (PAL) Default value (NTSC)
Menu “Screen Size
1
1. The parameters in the menu “Screen Size“ (Out H-centre, Out H-width, Out V-centre, Out
V-height) can vary from customer to customer, depending on the factory-set monitor settings im-
plemented by the manufacturer in the urodynamic system.
Out H-Centre 65 66
Out H-Width 81 80
Out V-Centre 160 130
Out V-Height 260 215
Menu “Manual Set
2
2. The parameters in the menu “Manual set“ (VGA left, VGA width, VGA top/4, VGA bott/4) can vary
from customer to customer, depending on the factory-set monitor settings implemented by the
manufacturer in the urodynamic system.
VGA Left 1 1
VGA Width 42 42
VGA Top/4 9 7
VGA Bot/4 266 266
Menu “Advanced
Video Standard PAL NTSC
Output Signal RGB RGB
H. Soften Off Off
Sense 1 1
RGB Term. Off Off
Sync In VGA HV VGA HV
Sync Out -CS-CS -CS-CS
Menu “Engineering
ADC Ref. 131 131
DAC Ref. 65 65
CV Filter 0 0
Adj. Osc. 0 0
SC/H Phase 4 4
Y/C Delay 1 1