Frequency Response Verification
This test checks if the amplitude flatness is within the spec range using a
Power Meter.
1. Turn on the generator and choose CH1 as the operating channel. Set the
Load to 50 Ω.
2. Connect the Power Meter, PC and generator as shown below.
3. Select Sine waveform of the generator and set the amplitude to 1 V and 2.5
V, frequency to the values listed below in sequence.