GD1 Page 34 of 76 850-816926-EN R06
19
Save the alignment settings to the system by
entering the following command
#SAV
Terminal response:
IM:OK
This saves the current settings,
exits Service Mode and initialize
(reboots) the system.
Note! After initializing has completed the GD1 will enter Measuring Mode (normal
operational mode). The initialization can take up to 5 minutes.
Note! As a default the zero filter is activated. The zero filter removes any fluctuations below
14 ppm*m (7% FS). So the analogue output will show 4 mA as long as the measurement is
below 14 ppm*m.
Fine tuning finished!