50 iGlucose®
Note: Your iGlucose® Blood Glucose Meter has been tested to ensure
that there is no change in the performance or external materials of
the device after 550 cleaning cycles and 550 disinfecting cycles. The
testing simulates 2 cleaning and disinfecting cycles per week over the
typical life of the meter (5 years).
To clean the iGlucose Lancing Device:
Thoroughly wipe the entire surface of the iGlucose Lancing Device
with disinfecting wipes listed above to clean any possible dirt, dust,
blood and other body fluids.
To disinfect the iGlucose Lancing Device:
1. Take another disinfecting wipe and wipe the iGlucose Lancing
Device thoroughly. (Note: All blood and body fluids should be
cleaned from surface before performing disinfecting procedure).
2. Allow the surface to remain wet for 2 minutes.
Note: Your iGlucose Lancing Device has been tested to ensure
that there is no change in the performance of the device after 550
cleaning cycles and 550 disinfecting cycles. The testing simulates
2 cleaning and disinfecting cycles per week over the typical life of
the meter (5 years).
2 3
Allow to remain
wet for 2 minute
1