2
2. That the device is operating and has been set up correctly:
a. date/time setting;
b. mains water hardness setting;
c. demineralised water yes/no setting;
d. setting of the presence of optionals as appropriate for the model concerned (e.g. P3, Acquastop, “FLSW” flow
switches, etc).
In addition to the above checks, the engineer must:
1. Run program “16 SERVICE”, preferably twice in succession, to prime the detergent intake lines correctly.
2. Run a complete cycle (with drying phase if featured in the model) and test the device to ensure that it is operating
correctly (e.g. test program accessible from the device's SETUP menu).
The operations summarised above must be performed and recorded in accordance with the checklist provided by the
manufacturer, using the “IN-SERVICE TEST RECORD” form (“REGISTRATION OF CORRECT OPERATION AT THE USER'S
SITE”) .
The authorised engineer is responsible for training the staff assigned to use the device at the time of installation.
The device may only be used by specifically trained staff.
The training of the staff responsible must include specific information on the possible risks involved in the use of the device, and
training in the safest possible way of carrying out the operating procedures.
The installation engineer is also responsible for notifying the responsible authority of the USER and SUPERUSER passwords for
access to the setup parameters.
2.1 TEST PROGRAM
For firmware version prior to 4.6: the test program consists of just 2 phases plus a short drying phase (only if the drying system is
installed on the model.
- PHASE 1: cold wash
- PHASE 2: thermal disinfection at 93 ° C, 10 minutes.
Water intake per phase: 9 litres (shown in the illustration as “Cold Water: 90”, referring to the value in decilitres. 90 decilitres = 9
litres).
Test program detail, for fw version prior to 4.6
===========================================================================
PHASE 1
Cold Water: 90
P1 DETERGENT (ml/l): 10
P3 DETERGENT (ml/l): 10
TARGET Temperature 1 (°C): 0 Extension (min): 1
=========================
PHASE 2
Demi Water: 90
P2 DETERGENT (ml/l): 10
TARGET Temperature 1 (°C): 93 Extension (min): 10
=========================
DRYING
Drying temperature (°C): 110
Drying duration (min): 10
From fw version 4.6: the test program consists of 2 phases plus a short drying phase.
The PHASE 1 has been modified: T_target 93°C, holding time 10 min. The principles for activation
of the dispenser pump and drying system (when installed) are unchanged.
- PHASE 1: thermal disinfection at 93 ° C, 10 minutes, chemicals dosage
- PHASE 2: thermal disinfection at 93 ° C, 10 minutes.
The purpose of using two successive PHASES at 93 ° C is to allow an eventual adjustment with external calibrated probe, during the
first phase, and to check the proper maintenance of the calibration in the second phase.