15
Steps Yes No Comments
Restore presets
31
Alarm volume: High
Compression rate: High
Continuous therapy: 80mmHg
Intermittent therapy: 80mmHg for 5 min
0mmHg for 2 min
Total Patient Therapy: 0
Component replacement and vacuum check
32
O-ring replaced with new o-ring.
33
Odor filter replaced with new odor filter.
34
The vacuum measured by the pressure gauge is
25 ± 5mmHg when device set @ 25mmHg.
35
The vacuum measured by the pressure gauge is
200 ± 5mmHg when device set @ 200mmHg.
36
Entered maintenance mode and reset annual timer New expiration date
Storage and battery maintenance
37
.%001 - %09 neewteb si level egrahc yrettaB
Battery fully charged.
Cleaning and Packing
38
Device and power supply cleaned and packaged
with all necessary items for customer use or return.
Performed by:
Print Name: Date:
Signature: Day Month Year